Fabrication of hemoglobin/ionic liquid modified carbon paste electrode based on the electrodeposition of gold nanoparticles/CdS quantum dots and its electrochemical application
- Release time:2024-08-09
- Hits:
-
-
-
Affiliation of Author(s):
化学与化工学院
-
-
Journal:
Electroanalysis
-
-
-
Key Words:
中文关键字:量子点,英文关键字:gold nanoparticles, quantum dots, ionic liquid,ele
-
Abstract:
A novel H2O2 amperometric biosensor,based on the electrodeposition of gold nanoparticles (AuNPs) and CdS quantum dots (CdS QDs) onto a carbon paste electrode (CPE) and immobilizing hemoglobin (Hb) with ionic liquid (IL), was presented in this article. The modified process of the electrode was monitored by scanning electron microscopy (SEM), cyclic voltammetry (CV) and electrochemical impedance spectra (EIS). Due to synergistic effects of AuNPs, CdS QDs and IL, the biosensor exhibited high stability and good bioelectrocatalytic ability to H2O2 with a linear range of the concentration from 10 to 750μM and detection limit of 4.35 μM (S/N=3).
-
Note:
董社英
-
-
First Author:
dongsheying
-
Indexed by:
Journal paper
-
-
-
-
-
-
Volume:
卷:24
-
Issue:
期:7
-
Page Number:
页:1554-1560
-
-
-
Translation or Not:
no
-
-
Date of Publication:
2012-07-01
-
-